Instrument introduction:arium®mini adopts secondary purification technology. The tap water connected to the first stage passes through the pretreatment column by means of the membrane pump. Pretreatment column includes activated carbon, refiner and downstream reverse osmosis combination to effectively remove particles, salt, chlorine and impurities. The pretreatment water stored in the water storage bag in the second stage is removed from the organic residues by ultraviolet lamp (185\254nm), and finally processed by pure water column, and the ultrapure water is obtained through sterile terminal filter (ASTM Type I).
Technical parameters:
Influent water quality |
Running water |
Water quality level |
ASTM I ultrapure water |
Reverse osmosis pure water |
Water production speed |
1L/min |
8L/h |
Electrical conductivity |
0.055μS/cm(25℃) |
<20μS/cm(25℃) |
Resistivity |
18.2MΩxcm(25℃) |
<0.05MΩxcm(25℃) |
TOC total organic carbon (with UV lamp) |
<5ppb |
- |
Microbial content |
<1cfu/ml |
<1cfu/ml |
Particulate matter content (>0.2 μm) |
<1/ml |
<1/ml |
Reverse osmosis ion removal rate |
- |
> 98% |
Reverse osmosis organic matter removal rate |
- |
> 99% |
Removal rate of reverse osmosis microorganisms and particles |
- |
> 99% |